Niti-S Hot SPAXUS™ Stent & Electrocautery Stent Delivery System





Lumen-apposing Metal Stent for the drainage of pancreatic pseudocyst or a gallbladder through a transgastric or transduodenal approach. Pre-loaded on the Electrocautery Delivery System.
Video clip
Indication
for the drainage of pancreatic pseudocyst or a
gallbladder through a transgastric or transduodenal approach
Features
Lumen-apposing
SPAXUS™ Stent
• Prevents
migration and maintains anastomotic conduit between two adjacent organs
• The
flexible flare design helps accommodative apposition regardless of the wall
thickness
• 8,
10, 16mm diameters enable to apply to various indications
Electrocautery-enabled
delivery system
• Select
the pad specified by generator manufacturer
• Set
to Pure Cut Mode 80-120 Watts, 400-500Vp
• Connect
the plug to the electrosurgical unit with a compatible active cable
Electrocautery-enabled
delivery system
• Step 1.
When the outer X-ray marker overlaps with the inner X-ray marker
→ The distal flange is completely
opened
• Step 2.
When the blue marker is visible under endoscopic view
→ Start opening proximal flange